INATEC – PTU-8012 Low Turbidity Sensor
INATEC – PTU-8012 widely used in such fields as sewage plants, Drinking water plants, water stations, surface water, and industries for inspecting the turbidity.
Characteristics
Range : 0.01 – 100 NTU.
Resolution : The deviation of reading in 0.001 ~ 40 NTU is ± 2 % or ± 0.015 NTU, choose the bigger one; and it is ± 5 % in the range of 40 – 100 NTU.
Flow Rate: 300 ml/min ≤ X ≤ 700 ml/min.
Pipe Fitting : Injection Port : 1/4 NPT, Discharge Outlet : 1/2 NPT.
Specification
| Information | Spesification |
|---|---|
| Product | Turbidity Sensor |
| Measure range | 0.01 - 100 NTU |
| Measurement accuracy | The deviation of reading in 0.001 ~ 40 NTU is ± 2 % or ± 0.015 NTU, choose the bigger one; and it is ±5% in the range of 40 - 100 NTU |
| Flow rate | 300 ml/min ≤ X ≤ 700 ml/min |
| Pipe fitting | Injection Port : 1/4 NPT, Discharge Outlet: 1/2 NPT |
| Environment temp | 0 ~ 45 ℃ |
| Calibration | Standard Solution Calibration, Water Sample Calibration, Zero Point Calibration |
| Cable length | Three-meter standard cable, it is not recommended to extend |
| Main materials | Main Body : ABS + SUS316 L |
| Sealing Element : Acrylonitrile Butadiene Rubber | |
| Cable : PVC | |
| Ingress protection | IP66 |
| Weight | 2.1 KG |